| AE50 | AE 100 | AE 160 | AE200 | |
| Readability | 0.1 mg | 0.1 mg | 0.1 mg | 0.1 mg |
| Massing range | 0-55g | 0-109g | 0-162g | 0-205g |
| Tare range (subtractive) | 0-55g | 0-109g | 0-162g | 0-205g |
| Reproducibility (standard deviation) | 0.1 mg | 0.1 mg | 0.1 mg | 0.1 mg |
| linearity | +-0.2mg | +-0.2mg | +-0.2mg | +-0.3mg |
| linearity relative to 10g | +-0.1mg | +-0.1mg | +-0.1mg | +-0.1mg |
| Stabilization time (typically) | 5 sec |
| Integration time (adjustable) | 1.5/3/6 sec |
Display sequence:
Mettler Delta Display off 0.4 sec
Mettler Delta Display on 0.2/0.4 sec
Stability detector:
Sensitivity selectable in three steps 1/2/off
Sensitivity drift (10-30deg.C):
+-2X10-6/deg.C
Calibration weight (built-in):
adjusted to an apparant mass of 8.0g/cc in an air density of 1200 mg/l 100g, adjustable to
+-0.1mg (50g, adjusted to +-0.1mg for AE 50)
Dimensions:
Weighing pan 80 mm dia.
Open space above pan 215 mm
Balance housing (WxDxH) 205x410x290 mm
Net weight 10.3 kg
Power supply:
Voltage, adjustable 115V, 220V
Admissible voltage range 92...132V, 184...265V
Frequency 50...60 Hz
Power consumption 10 VA
Admissible ambient conditions during operation:
Temperature 10...40 deg.C
Relative humidity (non-condensing) 25...85%