SPECIFICATIONS

  AE50 AE 100 AE 160 AE200
Readability 0.1 mg 0.1 mg 0.1 mg 0.1 mg
Massing range 0-55g 0-109g 0-162g 0-205g
Tare range (subtractive) 0-55g 0-109g 0-162g 0-205g
         
Reproducibility (standard deviation) 0.1 mg 0.1 mg 0.1 mg 0.1 mg
linearity +-0.2mg +-0.2mg +-0.2mg +-0.3mg
linearity relative to 10g +-0.1mg +-0.1mg +-0.1mg +-0.1mg

 

Stabilization time (typically) 5 sec
Integration time (adjustable) 1.5/3/6 sec

Display sequence:
Mettler Delta Display off 0.4 sec
Mettler Delta Display on 0.2/0.4 sec

Stability detector:
Sensitivity selectable in three steps 1/2/off

Sensitivity drift (10-30deg.C):
+-2X10-6/deg.C

Calibration weight (built-in):
adjusted to an apparant mass of 8.0g/cc in an air density of 1200 mg/l 100g, adjustable to +-0.1mg (50g, adjusted to +-0.1mg for AE 50)

Dimensions:
Weighing pan 80 mm dia.
Open space above pan 215 mm
Balance housing (WxDxH) 205x410x290 mm
Net weight 10.3 kg


Power supply:
Voltage, adjustable 115V, 220V
Admissible voltage range 92...132V, 184...265V
Frequency 50...60 Hz
Power consumption 10 VA

Admissible ambient conditions during operation:
Temperature 10...40 deg.C
Relative humidity (non-condensing) 25...85%