TROUBLESHOOTING

Problem Reason/(Correction)
Dark display? -no power to the instrument
-defective fuse
-OFF- appears in display? -a temporary power failure has occured
-(Press control bar)
A broken line appears in upper display? -the massing range has bee exceeded.
-calibration mass is activated.
-a mass was on the pan when the instrument was switched on.
A broken line appears in lower display? -pan is not installed.
-a mass was on the pan when the instrument was switched on.
Fluctuation in reading? -air currents. (Close balance door(s)).
-unstable foundation. -(Raise integration time.)
-the object being massed is not at room temperature.
Mass result is incorrect? -(Calibrate balance).
Only a portion of the display functions? -a temporary malfunction. (Pull out power cable and plug back in).
A blinking broken line appears in middle display? -unstable foundation and/or air currents. -(Close balance door(s) and/or raise integration time/stability settings).
-CAL Err- appears in display? -mass remained on the pan while entering calibration mode.  -(Press and hold control bar down. Returns to massing mode).
-CAL- does not appear in display? -a temporary malfunction. (Recalibrate balance).
A zero display does not appear after taring? -unstable foundation (Close balance door(s) and/or raise integration time/stability detector).